LightLas 532 Operator's Manual Rev. No 01 Page 22 of 115
Doctor Safety Filter : OD4 at 532 nm
Safety Class : Class 4
Power Display accuracy: Better than +/-20% of actual
Beam Divergence : < 0.2 NA
____________________________________________________________________
Aiming Laser
Laser Type : Red Laser Diode
Wavelength : 635650 nm (Red)
Mode of operation : Continuous Wave (CW)
Output power : Maximum of 1.0mW
Power adjustment : Continuously variable
Safety Class : Class 2
___________________________________________________________________
Laser Delivery Units (LDU)
Integrated Slitlamp LDU
Slitlamp model : CSO Model SL980 (Detailed refer to p.32)
Spot Sizes : 50, 100, 200, 300, 400, 500, 1000 m selectable
on Zoom assembly of Delivery unit
Focus plane : All spot sizes Parfocal to Slitlamp focus plane
Fiber Length : 2 m
Mounting : Direct to Slitlamp housing
Safety Filter : Fixed filter OD 4 @ 532nm
Beam Divergence : Cone angle of 20°
Attachment Slitlamp LDU
Slitlamp model : To attach on CSO Model SL990 (Detailed refer to p.32),
Haag Streit models BM or BQ, most Haag Streit Clone
Slitlamps and Zeiss SL30
Spot Sizes : 50, 125, 200, 300, 500m selectable on housing of
Delivery unit
Focus plane : All spot sizes Parfocal to Slitlamp focus plane
Fiber Length : 2 m
Mounting : On adapters fitted to Slitlamp Tonometer Mounts
Safety Filter : Fixed filter OD 4 @ 532nm
Beam Divergence : Cone angle of 20°
___________________________________________________________________
Laser Indirect Ophthalmoscope
Indirect model : Keeler All Pupil II
Retinal Spot size : 300µm nominal at focus with 20D Laser Lens
Illumination power : From Laser Console or stand alone power source
Fiber Length : 3 m