LightLas 532 Operator's Manual Rev. No 01 Page 97 of 115
Section 8 TROUBLESHOOTING
The Lightlas system 532 is designed to have a service free part as possible. It is
divided into three major sections such as, Symptom, Warning, and Error code.
8.1 Symptom
A completed listing of Lightlas system 532 symptoms or warnings and along with the
suggested corrective action is explained as following:
Symptom(s)
Cause(s)
Remedy action(s)
No power / no display
-Power failure
-Remote control panel
connector not
connected
-Check the power source or
connector
-Check the power fuses
-Check the emergency button is
outward ON position
-If the problem persists, call the
authorized service or
manufacturer for further
assistance
No lamp illumination
(SD only)
-No power to the lamp
-Lamp is burned out
-Check the lamp voltage
distribution (+12)
-Check the lamp integrity
-Check that there is no dust and/or
fingerprint on the bulb
-Call the authorized service or
agent for assistance
No aiming beam (SD
only)
-Aiming beam intensity
needs to be calibrated
-Aiming beam electric
circuit failure
-Reboot the system
-If the problem persists, call the
authorized service or
manufacturer for further
assistance