LightLas 532 Operator's Manual Rev. No 01 Page 30 of 115
All operating conditions and switches are shown on the LCD control panel. The LCD
control panel display consists of:
Laser Power
Laser Power Pulse Duration
Pulse Interval
Accumulated number of pulses
The type of LDU connected
Aiming intensity
Mode of operation
SP Mode Selection
Customized treatment configuration
Query or Help functionality
The mode of operation is an important function display because when the laser is
turned ON STANDBY mode is automatically selected, which prevents any accidental
firing of the LightLas 532 Laser System with LCD panel. In STANDBY mode, the
footswitch is disabled and the shutter module blocker will obstruct the beam path is
closed.
Only ‘Standby’ button switch is toggled, the system will be in treat mode. Then the
footswitch and shutter are now enabled and the aiming beam is turned on. If the
footswitch is pressed, the Green laser beam will be delivered into the fiber. The
system will turn back to STANDBY mode in the following situations:
No controls are operated for 10 minutes
Any warning or error condition occurs
Prior to activate the ‘Treat’ mode, it is recommended that all operating conditions are
to be set correctly such as patient positioning, power selection, pulses duration,
interval duration, aiming beam intensity, spot size and illumination intensity. This will
prevent the likelihood of accidental firing of the Green laser or unintentional delivery
during the set up stages.
Output power distribution can fine tune through an up/down arrow switch button on
the LCD touch control panel. The power setting will remain the same whenever the
power is on which means the default setting screen will be primary unless it is
replaced by another setting. The power can be adjusted from 50 to 2000 mWatts.
The pulse duration can be adjusted from 0.01 secs to 3.0 secs. by pressing the action
key switch buttons on the LCD touch control panel. Similarly, the Repeat Interval