LINXON Operation | 6
LX218-Operating-Manual-en1-04-(1908) 61 / 90
During a calibration with an external calibration leak or an internal calibration leak in
the "Manual internal" mode, the stability of the signal must be confirmed using the
"OK" button.
During a calibration with an external calibration leak, the following prompt appears:
Close calibration leak valve
When using tracer gas H
2
, wait 5 minutes
Confirm with OK
During a calibration with external tracer gas or internal tracer gas in the "Manual
internal" mode, the stability of the signal must be confirmed using the OK button.
After the end of the calibration process, the result is displayed.