LINXON Warning and error messages | 7
LX218-Operating-Manual-en1-04-(1908) 67 / 90
No. of the
error
message
Displayed message Description and possible remedy for the
cause
E35 24 V at OPTION output is
too high
The 24V for the external outputs I/O;
RS485; GAUGE HEAD is too high. (U >
30V)
Check the external feed of the 24V
outputs.
Power supply unit defective
E36 24 V at OPTION output is
too low
The 24V for the external outputs I/O;
RS485; GAUGE HEAD is too low. (U < 20V)
Fuse F1 on I/O board defective.
Power supply unit defective
E37 24 V power supply for the
remote control is too low (<
20 V)
The 24V for the external RC outputs and
fans 1+2 is too low. (U < 20V)
Fuse F2 on I/O board defective.
Power supply unit defective
E39 Temperature at electronic
unit is too high! (>60°C)
The ambient temperature is too high.
Unfavorable position of the leak detector.
(Heat build-up)
Fan failed.
Air filter too heavily contaminated.
Temperature sensor defective.
E41 TMP frequency is too low! The target speed (1450Hz) of the turbo
molecular pump (TMP) was not reached
within 5min.
Initial pressure of the TMP is too high.
Turbo molecular pump is defective.
Electronic drive unit TC 110 is defective.
E42 Emission off (p1 too high) Inrush of air
Valve V1 leaking.
E43 Emission off (p2 too high) Emission is switched off during normal
operation of the leak detector when the
pressure
in GROSS is p2 > (GROSS pressure
threshold + 5 mbar) or
in FINE is p2 > (FINE pressure threshold + 1
mbar) or