LINXON Warning and error messages | 7
LX218-Operating-Manual-en1-04-(1908) 71 / 90
No. of the
warning
message
Displayed message Description and possible remedy for the
cause
After confirmation of this warning message,
a warning triangle will remain in the "Ready"
menu, indicating the presence of this fault.
Only if this fault is eliminated will this
warning triangle disappear.
W104 Capillary broken • Capillary broken or torn
• Upper flow limit is set incorrectly
After confirmation of this warning message,
a warning triangle will remain in the "Ready"
menu, indicating the presence of this fault.
Only if this fault is eliminated will this
warning triangle disappear.
W105 Global Reset A global reset has been performed.
W106 Default settings loaded! The factory settings have been loaded by
the device software.
W107 Service interval has
expired!
• Perform service on the backing pump
• Perform service on the TMP
After confirmation of this warning message,
a warning triangle will remain in the "Ready"
menu, indicating the presence of this fault.
When maintenance is complete, the service
interval can be reset in the menu under
”Maintenance & service" (device -
maintenance interval).
W108 ZERO function is blocked. The ZERO function has been blocked in the
Settings menu, but activated through the
PLC input.
W109 Preamplifier signal is too
large (> 10 V)
The preamplifier signal is overmodulated in
the least sensitive measurement range.
• Preamplifier defective
• Mass spectrometer is heavily soiled
W120 Time-out during calibration Pressure threshold for following
measurement range is not reached within
the set time limit. The time limit is specified
by the set maximum evacuation time in the
configuration menu.
W121 The calibration leak signal
is too small!
• The calibration leak used for the
calibration is too small.