7 | Warning and error messages LINXON
72 / 90 LX218-Operating-Manual-en1-04-(1908)
No. of the
warning
message
Displayed message Description and possible remedy for the
cause
The external calibration leak valve is not
open or defective.
Internal calibration leak is defective
W122 Signal is unstable during
calibration!
Leak rate signal too small and noisy
Internal calibration leak is defective
Backing pump has highly unstable final
pressure
W123 Peak out of tuning range! Leak rate signal was unstable during the
mass adjustment. Re-calibrate.
Check internal calibration leak and repeat
the calibration with an external calibration
leak.
W124 Signal difference between
opened and closed
calibration leak is too small.
Internal calibration leak is defective.
The external calibration leak valve is
defective or not closed.
The calibration leak used for the
calibration is too small.
W125 Calibration factor too low The calculated calibration factor is outside
the permissible range (< 0.1). The old factor
is retained.
The calibration leak is defective.
The leak rate value entered for the
calibration leak is much too small.
The conditions necessary for calibration
have not been satisfied.
W126 Calibration factor too high! The calculated calibration factor is outside
the permissible range (> 100). The old factor
is retained.
The calibration leak is defective or empty.
The calibration leak value entered for the
calibration leak is too large.
Mass spectrometer soiled and non-
sensitive.
The conditions necessary for calibration
have not been satisfied.