MODEL 9DP-1 Operator’s Manual
Ludlum Measurements, Inc. 32 December 2015
Operating Mode
“Operating Mode” represents the current operating mode for the data logging
feature. “Operating Mode” is a special field with valid values of “Off” or “On.”
Delay On Power Up
“Delay On Power” represents the amount of time in seconds that the instrument
must wait on power-up before data logging starts logging data. “Delay On Power
Up” is a numeric field with valid values from 1 to 180.
Sample Period
“Sample Period” represents the amount of time in seconds that the instrument
must wait between reporting data logging events. “Sample Period” is a numeric
field with valid values from 1 to 3600 (1 hour).
Report Setup
“Report Setup” is a sub-menu which grants access to the “System,”
“Instrument,” and “Readings” values that are to be reported by the data logging
feature.
System
“System” is a sub-menu which grants access to the “Date,” “Time,” and
instrument settings to be reported by the data logging feature.
Date
Enabling “Date” will cause the data logging feature to report the current
date during each logging event. Otherwise, the date field should be
disabled. “Date” is a special field with valid values of “Off” and “On.”
Time
Enabling “Time” will cause the data logging feature to report the current
time during each logging event. Otherwise, the time field should be
disabled. “Time” is a special field with valid values of “Off” and “On.”
Chamber Temperature
Enabling “Chamber Temperature” will cause the data logging feature to
report the temperature currently being recorded by the ion chamber
during each logging event. Otherwise, the chamber temperature field
should be disabled. “Chamber Temperature” is a special field with valid
values of “Off” and “On.”