MODEL 9DP-1 Operator’s Manual
Ludlum Measurements, Inc. 34 December 2015
Readings
“Readings” is a sub-menu which grants access to the instrument readings to
be reported by the data logging feature. This includes both the rate and
integrated dose readings.
Reading
Enabling “Reading” will cause the data logging feature to report the
current rate reading during normal logging events. The value of
“Reading” can also report other values depending on the event type. For
instance, if reporting the integrated dose (Event Type = 2) on power-
down is enabled, “Reading” will represent the integrated dose value if
enabled. If the reading is not to be reported, the field should be disabled.
“Reading” is a special field with valid values of “Off” and “On.”
Reading Multiplier
Enabling “Reading Multiplier” will cause the data logging feature to
report the current rate reading multiplier during normal logging events.
The value of “Reading Multiplier” can also report other values depending
on the event type. For instance, if reporting the integrated dose (Event
Type = 2) on power-down is enabled, “Reading Multiplier” will represent
the integrated dose multiplier if enabled. If the reading multiplier is not to
be reported, the field should be disabled. “Reading Multiplier” is a special
field with valid values of “Off” and “On.”
Reading Units
Enabling “Reading Units” will cause the data logging feature to report
the current rate reading units during normal logging events. The value of
“Reading Units” can also report other values depending on the event
type. For instance, if reporting the integrated dose (Event Type = 2) on
power-down is enabled, “Reading Units” will represent the integrated
dose units if enabled. If the reading units are not to be reported, the field
should be disabled. “Reading Units” is a special field with valid values of
“Off” and “On.”
Integrated Reading
Enabling “Integrated Reading” will cause the data logging feature to
report the current integrated dose reading during normal logging events.
Otherwise, the field should be disabled. “Integrated Reading” is a special
field with valid values of “Off” and “On.”