104
57-606 ECLIPSE Model 706 Guided Wave Radar Transmitter
A warning screen is shown that the loop can be returned to
automatic control.
At this point the echo rejection curve can be viewed by
selecting Rejection Curve as Curve 2 in the lower left corner
of the Echo Curve screen. The Rejection curve will then be
displayed in red as shown in the screenshot below.
4.4 Buildup Detection
A unique feature contained within the Model 706 can be
used to obtain an indication of build-up along the length of
the probe. This can be set as the HART SV or TV which
can be monitored in the control room. An algorithm com-
pares the buildup echo strength as compared to the Lvl
Thrsh Value, and outputs value in percent.