211 
 
The integral calculation will be stopped when the deviation of the reference and feedback is greater than 
  the setting. This function is used for reducing overshoot and oscillation incurred by the integral action in 
the dynamic process.   
 
Output values used for limiting the integral adjustment:   
When the PID output serves as the frequency reference, use the maximum output frequency as 100%.   
When the PID output serves as the torque reference, use 300% rated torque as 100%.   
 
0: Perform differential control for the deviation of reference value and feedback value 
1: Perform differential control for feedback value, applicable to frequent and quick changes of the 
reference value.   
 
Output values used for limiting the differential adjustment. 
When the PID output serves as the frequency reference, use the maximum output frequency as 100%.   
When the PID output serves as the torque reference, use 300% rated torque as 100%.   
 
This is not only the sampling cycle for the feedback, but also the control cycle for PID. The PID regulator 
calculates once for each sampling cycle. The bigger the sampling cycle is, the slower the response 
becomes.  
 
When the deviation of the reference value and feedback value relative to the percentage of reference 
value is less than the deviation limit setting, the PID will stop adjusting and the output remains unchanged. 
This function can avoid frequent actions of the PID control.   
 
Fig. 6-73      Schematic diagram of deviation limit 
 
PID u
er limit channel 0~3
0
P14.14 
Deviation limit 0.0~20.0%
2.0%
P14.13 
Sam
lin
 c
cle 0.001~50.000s 
0.010s
 P14.12 
Differential am
litude limit 0.0~100.0%
10.0%
P14.11 
Differential control selection 0~1 
0
P14.10 
Inte
ral am
litude limit 0.0~100.0%
100.0%
P14.09 
Inte
ral se
aration threshol
0.0~100.0%
30.0%
P14.08