30 D33451 | Date 01/2018
Operation and control
4. Now, as described under 5.5.2 , set the individual values which you want the appliance
to operate at. The set values can be changed at any time while the timer elapses. The
changes are effective immediately.
In
Setup, you can choose if the timer should run setpoint-dependent or not. This deter-
mines whether the timer should not start until a tolerance band around the set tempera-
ture is reached or if it should start right after activation (see page 50). If the timer runs
setpoint-dependent, this is indicated by the
symbol in the timer display.
When the timer expires, the display shows 00h:00m.
All functions (heating etc.) are switched off. In addition, an
acoustic alarm sounds, which can be turned off by pressing the
confirmation key.
Note that condensation may occur in the interior after you
switch off the heating.
To deactivate the timer, open the timer display by pressing the
activation key again and then turning the turn control to reduce
the timer setting until --:-- is displayed. Confirm with the confirma-
tion key.
5.5.6 Programme mode
In this operating mode, programmes saved in the appliance can be started with different
combinations of individual parameters (temperature, humidity, etc.) at staggered intervals,
which the appliance then automatically processes in sequence. These programmes are not cre-
ated directly at the appliance but externally at a computer / laptop and using AtmoCONTROL
software. Transfer to the appliance is possible using the provided USB storage medium or via
Ethernet.
A description of how to create and save programmes can be found in the separate
AtmoCONTROL software manual.
NOTICE
One or several default sterilisation programmes are saved in the appliance (see page 58).
They only serve to sterilise the appliance itself. Do not use them to sterilise medical devices.
The hold time in the sterilisation programme set in the appliance at 180 °C is 1 h. The total
time including heating up and cooling down to 50 °C is 6 h 30 min.
Starting a programme
1. Press the activation key to the right of
the status display. The current operating
mode is highlighted automatically, in this
example
Manual mode ( ).
manueller Betrieb
Fr 20.10.2010 20:31
80
GRAPH
8
0
GRAPH
APH
Manual mode
17:4413.Sept.2012
Activate
2. Turn the turn control until the
start
symbol is highlighted. The current
programme is displayed, in this example
Steri 180.
Steri 180
ready
10:4412.Sept.2012
TIMER
00m00h
End
13:30 23.11.
TIMER
--m--h
End
9:00 23.11.