D33451 | Date 01/2018 35
Operation and control
5. With the turn control, select ON (
) or
OFF ().
ALARM
max
38.5°C
0.0K
min
35.5°C
auto
+
-
6. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
ALARM
K
min
°C
auto
max
38.C
35 5.
0.0
+
-
7. With the turn control, adjust the desired
tolerance band. We recommend 0.5 to
1 K.
ALARM
K
min
°C
auto
max
38.C
35 5.
1.0
+
-
8. Press the confirmation key to confirm.
Temperature monitoring is now active.
ALARM
max
38.5°C
1.0K
min
36.5°C
auto
+
-
5.6.2 Humidity monitoring
(only for appliances in the corresponding configuration)
If humidity monitoring was triggered, this
is indicated by the humidity display: the
actual humidity is highlighted in red and a
warning symbol
is shown ( Fig. 21 ). If the
acoustic alarm has been activated in menu
mode (
Sound, see page 59, as indicated
by the speaker symbol
), the alarm is
additionally signalled by an intermittent
acoustic signal. Information on what to do
in this case is provided in chapter Malfunc-
tions, warning and error messages from
page 46.
4
m
min
000
°
O
LIG
ON
ON
ON
ON
LIG
IG
IG
O
LIG
FEUCHTE
30.0%rh
Set 30.0%rh
0
°
C
°C
°C22.4
E
MP
°CSet 37.0
a
m
0
m
3
:30
HUMIDITY
Set 70
.0%rh
%rh
75.4
m
FEUCHTE
30.0%rh
Set 30.0%rh
C
°CSet 37.0
m
0
HUMIDITY
Set 70
.0%rh
%rh
75.4
Fig. 21
Humidity monitoring triggered