4-38 | Model 933 UV Analyzer
Cell Temp. Default
The default temperature for the Measuring Cell (°C or °F). This value
is used when Measured temperature compensation is disabled.
Fixed The analyzer will use the default value.
Measured The analyzer will actively measure the tem-
perature reading.
Cell Pressure (optional)
Fixed The analyzer will use the default value.
Measured The analyzer will actively measure the abso-
lute pressure reading. See “0–100 PSIA (0–700
kPag), 0–30 PSIA (0–210 kPag), or , 0–500 PSIA
(0–3450 kPag) Pressure Transducer”charts
below for jumper selectable pressure ranges.
Default
The default absolute pressure for the Measuring Cell (mmHg or
"Hg). This value is used in place of the measured Measuring Cell
pressure when active pressure compensation is disabled.
LowScale
The pressure transducer low value (mmHg or "Hg) corresponding
to 0.0 V
DC
. See “0–100 PSIA Pressure Transducer” chart below.
HighScale
The pressure transducer high value (mmHg or "Hg) correspond-
ing to 5.0 V
DC
. See “0–100 PSIA Pressure Transducer” chart below.
0–100 PSIA Pressure Transducer Chart:
Jumper Settings, Customer I/O Board
PSIA mmHg JP301 JP302
0–100 0–5171.5 A A
0–33 0–1706.6 A B
33–66 1706.6–3413.2 B B
66–100 3413.2–5171.5 C B