Controller / User Interface | 4-39
While the 0–30 PSIA pressure transducer can not be re-scaled, its
jumper settings are listed in the “0–30 PSIA Pressure Transducer”
chart.
0–30 PSIA Pressure Transducer Chart:
Jumper Settings, Customer I/O Board
PSIA Pressure Range JP301 JP302
0–30 0–1552 mmHg A A
If changing the jumpers, be sure to make corresponding changes in
the System 200 Configurator Software.
If the 0–500 PSIA pressure transducer is being used, the jumpers
(JP301/JP302) on the Customer I/O board allow you to re-scale the
pressure range. The maximum allowable Cell Pressure is 200 psig
(1380 kPag). See “0–500 PSIA Pressure Transducer” chart below.
0–500 PSIA Pressure Transducer Chart (High Pressure/Low Range Kit):
Jumper Settings, Customer I/O Board
PSIA Pressure Range JP301 JP302
0–166.667 0–8619.16 mmHg A B