4 Operation
761 Compact IC
158
Started Indication whether the sample has been started
or not:
0 sample not started
1 sample started
2…n
number of injections done
Level Calibration level (see section 4.4.5) for the
sample: Level 0 stays for normal analysis, levels
1 and greater for calibration runs. Correct filling
of this column enables to calculate calibration
coefficients automatically.
Volume Injected volume of sample in µL.
Amount Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
= C
c
• Amount
s
/ Amount
c
Shut down system after the queue finishes
If this option is enabled, the system hardware is
shut down after the sample queue has been fin-
ished (all pumps are switched off), but it re-
mains connected.
<Start> Start execution of the sample queue from the
first line with Started = 0.
<Pause> Stop execution of the sample queue after the
current determination has been finished.
<Edit> Open the sample queue editor program for ed-
iting the sample queue table (see section
4.6.3).
<Close> Close the sample queue overview table.
Start sample queue
SAMPLE QUEUE OVERVIEW / Control / Start
Start execution of the sample queue from the first row with
Started = 0.
For each row, the sample-specific data are transferred to the method of
the current determination and the parameter
Started is set to 1 for this
row.
If processing the sample queue is interrupted with
<Pause> then it can
be continued again with
<Start> (with a running chromatogram only
when data acquisition is finished).