4.3 Systems
761 Compact IC
81
Ident User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the
chromatogram window and in the
Chromatogram
open window.
Calibration level Calibration level (0 = sample; 1n = calibra-
tion solutions).
Info 1 / Info 2 Sample description.
Volume Injected volume in µL.
Dilution Dilution of the sample.
Vial number Autosampler vial position to take sample from.
Amount Sample amount. If this value is different for the
calibration run (c) and the sample run (s), the
component concentrations of the sample are
calculated as follows:
C
s
= C
c
Amount
s
/ Amount
c
Internal standard amount Concentration of the internal standard compo-
nent for relative concentration calculations.
Date/time when... Date and time of sample collection (the default
values are equal to the date and time when the
chromatogram starts).
The
Verify sample option is disabled if determinations are made with an
active sample queue.