5 Operation
■■■■■■■■
21
Setting Meaning Selection Default value
P1.7
SET DUE CAL
pH calibration time NO | HOURS | DAYS NO
P1.8
RESET SETTINGS
Reset pH settings YES | NO NO
P1.9
TEMPERATURE CAL
pH temperature offset YES | NO
P1.1 Buffer selection
P1.1 is used to select the buffer family for the electrode calibration.
1 to 3 points can be calibrated.
Press the
button during calibration to exit the calibration and save
the points calibrated up to that point.
The 949 pH Meter automatically detects (see chapter 5.5.3, page 24) 3
buffer families (METROHM, USA, NIST). A manual calibration (see
chapter 5.5.4, page 27) (CUSTOM) with up to 2 calibration points can
also be carried out with user-defined values.
METROHM buffer: 4.00– 7.00
1)
– 9.00
USA buffer: 1.68– 4.01– 7.00
1)
– 10.01– 12.45
NIST buffer: 1.68– 4.00– 6.86
1)
– 9.18– 12.46
Default value: METROHM
1)
Neutral point is always requested first.
The bottom left of the display shows a row of beakers in measuring
mode. This row shows which buffers were used for the last automatic cali-
bration and the last manual calibration.
P1.2 Resolution
P1.2 is used to select the resolution for the pH measurement.
0.1
0.01
Default value: 0.01
P1.3 Stability
P1.3 is used to select the stability of the pH measurement.
Metrohm recommends waiting for the stability of the measurement indi-
cated by the
symbol. The symbol will appear on the display if the
measurement is not stable.