5.5 pH parameters
■■■■■■■■■■■■■■■■■■■■■■
22
■■■■■■■■
LOW: The symbol is also shown if the measurement stability is
poor. The measured values are within 1.2 mV.
MEDIUM: Measured values are within 0.6 mV.
HIGH: The symbol is shown only if measurement stability is high.
The measured values are within 0.3 mV.
Default value: MEDIUM
P1.6 pH calibration values
P1.6 is used to display or print the results of the last calibration per-
formed.
VIEW
The following results are shown automatically on the display, one after the
other:
Date and time of calibration, beaker with buffers used
Offset value of electrodes in mV
1st slope in % in the measuring range
2nd slope in % in the measuring range (is shown only for 3 calibration
points)
NOTE
The 949 pH Meter accepts only calibrations with pH electrodes with a
slope in an acceptance range between 80% and 120%. Calibration can-
not be exits outside the acceptance range. The SLOPE OUT OF
RANGE error message appears on the display.
PRINT
The following results are printed out:
Model number and serial number
Date and time of calibration
Offset value of electrodes in mV
pH range with relative slope
NOTE
Use only the original printer specified by the manufacturer.
Ensure that the printer is connected and switched on.
Ensure that the paper roll and the cassette are inserted correctly.
Additional information can be found in the manual for the printer.