35
Item Edit:
ITEM: Set the test item. The following parameters are available at the right side:
OFF : Disable this measurement function.
Ls : Set this item as measure Ls (Serial Inductance).
Lp : Set this item as measure Lp (Parallel Inductance).
Cs : Set this item as measure Cs (Serial Capacitance).
Cp : Set this item as measure Cp (Parallel Capacitance).
Q : Set this item as measure Q (Quality Factor).
D : Set this item as measure D (Dissipation Factor).
Rs : Set this item as measure Rs (Serial Resistance).
Rp : Set this item as measure Rp (Parallel Resistance).
Z : Set this item as measure Z (Impedance).
θ : Set this item as measure θ (Phase Angle).
X : Set this item as measure X (Reactance).
Y : Set this item as measure Y (Admittance).
G : Set this item as measure G (Conductance).
B : Set this item as measure B (Susceptance).
Rdc : Set this item as measure Rdc (Direct Current Resistance).
VOLT: Set measurement voltage. The setting range is 10mV-2.0V.
FREQ: Set theMeasure frequency; this item can only be set under AC mode.
Setting Range:
10Hz~20KHz (6363)10Hz~100KHz (6364)
10Hz~200KHz (6365)0.1Hz~200KHz (6365A)
10Hz~500KHz (6366)0.1Hz~500KHz (6366A)
10Hz~1MHz (6367)0.1Hz~1MHz (6367A)
SPEED: Set theMeasure speed; there are four options available: MAX (fastest)/FAST
(fast)/MED (medium speed)/SLOW (slow).
MODE: Set the compare mode.
Abs: Its upper and lower limits are determined by the value entered in the
HI/LO fields.
ΔAbs: Its upper limit is STD+HI; lower limit is STD+LO.
Δ%: Its upper limit is STD + (STD x HI%); lower limit is STD + (STD x LO%).
STD: Set the standard value. The setting range is -999G - +999G.