VetPro DC Installation/Service Manual, 00-02-1606, Revision R01
120
Specifications
This chart contains information required to be provided to the user by 21 CFR.
Parameter
AC 110 V to 230 V, 50 Hz or 60 Hz
250 V, UL Recognized –branch circuits must not exceed 15A
5 amp
60 kV, 65 kV or 70 kV
5% of selected value
4 mA to 7 mA
±1 mA
20 ms through 2 s
Irradiation Time Accuracy
1 ms + 5% of indicated value
Minimum Source to Skin
Distance
8 inch (20 cm)
12 inch (30 cm)
Minimum Inherent Filtration
2 mm Al equivalent @ 70 kV
0.4 mm (per IEC 60336)
Automatic wait time is 15x the exposure time.
12.5 degrees
+10 °C to +35 °C (+50 °F to +95 °F)
-35 °C to +66 °C (-31 °F to +151 °F)
0 °C to +50 °C (+32 °F to +122 °F)
70 kPa to 106 kPa
70 kPa to 106 kPa
70 kPa to 106 kPa
10 to 80% non-condensing
10 to 80% non-condensing
3,000 m (9,843 ft.)
8 inch (20 cm) for bisecting angle technique
12 inch (30 cm) for paralleling technique
Diameter of 7 cm (2.72 in.) at the end of the 7-inch cone.
Cones with smaller diameter and rectangular beams are available.
D470237, D469182, D470589, and 6,837,468