VetPro DC Installation/Service Manual, 00-02-1606, Revision R01
59
System Function Checklist
Component
Directions
Wall Mounting
Verify that the wall support is adequate and that the system is properly mounted
to the wall.
Labels
Verify that all certified components bear labels that include the model and serial
number, date of manufacture and a statement of certification as noted elsewhere
in this manual.
Tubehead
Check for oil leaks or other evidence that could indicate internal damage. If
necessary, replace the Tubehead.
Tubehead
Rotation
Ensure that the Tubehead maintains its position around the horizontal axis while
remaining easy to rotate and position. Also check the vertical pivot of the
Tubehead for easy movement while remaining in position after moving.
Suspension
Check that all movements are smooth and quiet. Verify that the Tubehead is
properly counterbalanced for vertical drift and that the Horizontal and Articulated
Arms do not drift horizontally.
Power Switch
Verify that the switch is working properly and that the Ready Indicator is
illuminated when the power switch is in the ON position.
Operator Panel
Controls
With the power switch in the ON position, verify that technique factors appear on
the Operator Panel. Also, check the function of the selection buttons for Tooth
Selection, Image Receptor Type, and Patient Size. Pressing a selection button
will cause indicator lamps to indicate the selected item.
Exposure
Button
Verify that the Exposure button on the Operator Panel functions properly. To
make an exposure, press and hold the Exposure button until the Radiation
Indicator goes off and the audible signal stops.
Exposure
Indicators
Make several exposures and verify that the Radiation Indicator illuminates and
the audible signal is heard.
Premature
Termination
Select the longest exposure time possible using the up and down arrows. Initiate
an exposure but release the Exposure button after a brief period of time (before
the timer terminates the exposure). Verify that the display indicates “Pre-
termination Error” and returns to normal operating mode.
Coil-cord Hand
Switch Option
If a coil-cord handswitch is used, inspect the switch housing and coil cord for
damage or wear. Replace them if there is any evidence of damage.
User
Information
Verify that the owner of the system has received the user manual.