VetPro DC Installation/Service Manual, 00-02-1606, Revision R01
15
• Mobile RF communications equipment can affect the
performance of medical electrical equipment. Midmark
advises against using the VetPro DC system adjacent to other
devices. If it must be used near other devices, carefully adjust
their configuration to ensure that electromagnetic interference
(EMI) does not degrade performance. Test both devices for
normal operation.
• Use of accessories, transducers, or cables other than those
provided by or specified by Midmark can result in increased
electromagnetic emissions or decreased electromagnetic
immunity of this equipment. This can result in improper
operation.
• Use portable RF communications equipment (including
peripherals such as antenna cables and external antennas,
including cables specified by Midmark) no closer than 30 cm
(12 in.) to any part of the VetPro DC system. Otherwise,
performance of the equipment may be degraded.