VetPro DC Installation/Service Manual, 00-02-1606, Revision R01
81
X-ray Images
1. Adjust the selected exposure time, kilovoltage or tube current to
produce an acceptable image. If necessary, reprogram the techniques
factors, as explained in the System Configuration section of this manual.
2. Verify the kilovoltage and tube current during an exposure using the
diagnostic mode, as explained in the System Configuration section of
this manual. Alternatively, you may employ a non-invasive meter to
evaluate kilovoltage and exposure time.
3. Inspect the condition of the remaining imaging chain components such
as the film, chemistry and processor, or the condition of the X-ray
sensor and computer.
4. Check for pulsing output with fluorescent screen. If this is occurring,
replace the Logic Board.
5. Light image can be caused by the Logic Board. This is characterized by
a short pulse of exposure as viewed with a fluorescent screen. Replace
the Logic Board.
6. Full fluorescence with light image may require adjustment of the
preprogrammed techniques.
Operator Display
Stops at “VetPro
®
DC”
Replace the operator display board.
Continuous Beep,
Backlit Display,
No Display
Characters, No
Ready Light,
Power Supplies
OK
Replace the Operator Display Board [30-08054] or the Logic Board [30-
08160], or both.
Continuous Beep
with Normal
Display
Replace the Logic Board [30-08160].
Buzzing Noise is
Heard
Replace the Logic Board [30-08160].