32 Milli-Q® Reference System
Registering UV Lamp timer
The timer used for the UV 185 nm Lamp must be reset when the System is installed.
If this is not done, then the message indicating that a Lamp replacement is needed is
shown too early.
Before doing this, make sure that the date and time have been checked for accuracy.
This procedure shows how to reset the timer used for the UV 185 nm Lamp.
Place the System in STANDBY
Mode.
Press .
Press .
Select Install UV 185 nm Lamp.
Press .
Continued on next page