34 Milli-Q® Reference System
Calibrating the Flowrate
Introduction
The Milli- Water flowrate should be calibrated when the System is installed.
A 1 Litre graduated cylinder is needed.
Procedure
Follow the steps below to perform a Flow Calibration.
Step
Action
Diagram
1
Go to STANDBY Mode.
2
Select Menu.
Press .
3
Enter the Manager Menu.
See the Software Chapter to
learn how to enter the Manager
Menu.
4
Select Setup.
Press .
5
Select Flow Calibration.
Press .
Continued on next page