Milli-Q® Reference System 35
Calibrating the Flowrate, Continued
(continued)
Place a 1 L Graduated Cylinder
under the POD Unit.
Press .
Press .
Water dispenses automatically
from the POD Unit.
Wait until it stops dispensing
water.
Measure the amount of water
(in ml) that was dispensed.
Suppose 870 ml was collected.
Input this using the Keypad.
Repeat the flow calibration to
improve accuracy.
Press .
Press 3 times on .