Milli-Q® Reference System 25
Installing the Q-Gard® Pack
Procedure
Follow the steps below to install a new Q-Gar Pack.
Step
Action
Diagram
1
Start in STANDBY Mode.
NOTE:
The Q-GARD® PACK OUT Alarm
message is not shown at this
time. By following the
instructions earlier in this
manual, the alarm was
cancelled.
2
Open the left door of the
System Cabinet.
Remove the 2 protective caps
located on the ports inside.
3
Remove the covers on the 2
ports of the Q-Gard® Pack.
Make sure the rubber O-rings
are firmly in place.
Wet the O-rings with water.
4
Push the top of the Q-Gard®
Pack into the ports on the
System.
Continued on next page