Milli-Q® Reference System 61
Replacing the Q-Gar Pack
When
The Q-Gar Pack should be replaced when one of the following Alarm or Alert
messages is displayed.
Alarm message = MILLI-Q RES < SP, REPLACE Q-GAR AND QUANTUM®
Alert message = REPLACE Q-GARD® PACK
Removing
Remove the used Q-Gard® Pack by following the steps below.
Step
Action
Diagram
1
Place the system into STANDBY
Mode.
2
Push the POD Plunger down
once to depressurise the System.
After water stops being
dispensed, push down the
POD Plunger again.
3
Open the System left door.
Lift up the Pack Locking Handle.
Continued on next page