Milli-Q® Reference System 61
Replacing the Q-Gard® Pack
The Q-Gard® Pack should be replaced when one of the following Alarm or Alert
messages is displayed.
• Alarm message = MILLI-Q RES < SP, REPLACE Q-GARD® AND QUANTUM®
•
Alert message = REPLACE Q-GARD® PACK
Remove the used Q-Gard® Pack by following the steps below.
Place the system into STANDBY
Mode.
Push the POD Plunger down
once to depressurise the System.
After water stops being
dispensed, push down the
POD Plunger again.
Open the System left door.
Lift up the Pack Locking Handle.
Continued on next page