64 Milli-Q® Reference System
Replacing the Quantum® Cartridge
The Quantum® Cartridge should be replaced when one of the following Alert or Alarm
messages is displayed.
• Alert message = REPLACE QUANTUM® CARTRIDGE
• Alarm message = MILLI-Q RES < SP, REPLACE Q-GARD®
AND QUANTUM®
The Quantum® Cartridge should be replaced whenever the Q-Gard® Pack is replaced.
Follow the steps below to remove the used Quantum® Cartridge.
Place the System into STANDBY
Mode.
Push the POD Plunger down
once to depressurise the System.
After water stops being
dispensed, push down the
POD Plunger again.
Open the System right door.
Remove the used Quantum®
Cartridge.
In a few moments, the System
indicates that the Quantum®
Cartridge is removed.
Continued on next page