64 Milli-Q® Reference System
Replacing the Quantum® Cartridge
When
The Quantum® Cartridge should be replaced when one of the following Alert or Alarm
messages is displayed.
Alert message = REPLACE QUANTUM® CARTRIDGE
Alarm message = MILLI-Q RES < SP, REPLACE Q-GARD®
AND QUANTUM®
The Quantum® Cartridge should be replaced whenever the Q-Gar Pack is replaced.
Removing
Follow the steps below to remove the used Quantu Cartridge.
Step
Action
Diagram
1
Place the System into STANDBY
Mode.
2
Push the POD Plunger down
once to depressurise the System.
After water stops being
dispensed, push down the
POD Plunger again.
3
Open the System right door.
Remove the used Quantu
Cartridge.
4
In a few moments, the System
indicates that the Quantu
Cartridge is removed.
Continued on next page