24 Milli-Q® Reference System
POD Unit, tubing and power cord, Continued
Alarm messages
Because the System is starting without a Q-Gard® Pack or a Quantum® Cartridge
installed, there are alarm messages displayed.
These alarms are:
Q-GARD® PACK OUT and
QUANTUM® CARTRIDGE OUT.
NOTE:
The TANK EMPTY Alarm message is shown if the System is configured to have a Level
Sensor.
Cancel Alarms
When an Alarm message is displayed, follow the instructions on the screen to cancel
the text display of the Alarm.
Check the date
When the Alarm messages are cancelled, check that the displayed date is correct.
If necessary, go to the Manager Menu Software and correct the date and time. See the
Software Map in the beginning of the Software Chapter for more information.
Do not install a Q-Gard® Pack or a Quantum® Cartridge until the displayed date
is correct.