Milli-Q® Reference System 81
Summary of Alarm messages, Continued
Alarm messages
(continued)
LCD message
What it means
MILLI-Q RES > MAX
The Milli- Water resistivity is out
of measurement range.
Contact Millipore SAS.
MILLI-Q T < MIN
The Milli- Water temperature is
out of measurement range.
Contact Millipore SAS.
MILLI-Q T > MAX
The Milli- Water temperature is
out of measurement range.
Contact Millipore SAS.
POD LOCKED
The POD Unit microswitch is locked.
Push the Plunger all the way down
and release.
Q-GARD® PACK OUT
The Q-Gar Pack is not installed
correctly or it has been removed.
The System stops operating.
Verify that the Q-Gar Pack is
installed correctly.
Contact Millipore SAS if the problem
continues.
QUANTUM® CARTRIDGE OUT
The Quantum® Cartridge is not
installed correctly or it has been
removed. The System stops
operating.
Verify that the Quantum® Cartridge
is installed correctly.
Contact Millipore SAS if the problem
continues.
TANK EMPTY
The System has detected an empty
Reservoir.
Refill the Reservoir.
Verify that the Reservoir level sensor
is plugged into the System Cabinet.
WATER DETECTED
A Water Sensor (an accessory
connected to the System) has
detected water. The System stops
operating.
Clean up the spilled water.
Make sure the source of the leak is
fixed.