Page 70 Accumass BW500 PL-565
Parameters
P360 Calibration Duration
Sets the number of whole belt revolutions to be used during a zero or span
calibration. (f=1)
Enter number of belt revolutions, range 1 to – 99
P365 Multispan
Select the span reference to be applied for determination of rate and totalization.
Entry:
1 = multispan 1 (MS1), for product or condition A
f
2 = multispan 2 (MS2), for product or condition B
3 = multispan 3 (MS3), for product or condition C
4 = multispan 4 (MS4), for product or condition D
5 = multispan 5 (MS5), for product or condition E
6 = multispan 6 (MS6), for product or condition F
7 = multispan 7 (MS7), for product or condition G
8 = multispan 8 (MS8), for product or condition H
Refer to Multispan on page 50 and P270, Auxiliary Input Function (6) on page 67.
P367 Direct Zero Entry
Directly enters the zero reference count
Direct entry is intended for use when replacing software or hardware and it is not
convenient to perform an initial zero at that time. (f=0)
Refer to Recalibration on page 41 for execution.
P368 Direct Span Entry
Directly enters the span reference count for the span selected, span 1 to 8 (P368-01
to -08).
Direct entry is intended for use when replacing software or hardware and it is not
convenient to perform an initial span at that time. (f=0)
Refer to Recalibration on page 41 for execution.
P370 Zero Limit Deviation %
Sets the zero calibration deviation limit (±) from the last initial zero. If the
accumulated deviation of successive zero calibrations exceeds the limit, the zero
calibration is aborted. (f=12.5)
Note:
If the ‘legal for trade’ certification switch has been set, the zero limit is ±2%.
Enter the maximum allowable % deviation