PL-565 Accumass BW500 Page 85
Parameters
Test and Diagnostic (P900 - P951)
These parameters are used for test and diagnostic purposes.
P900 Software Revision
Displays the EPROM (Flash ROM) software revision level.
P901 Memory Test
Tests the memory. Test is initiated by scrolling to the parameter or repeated by
'pressing enter'
Display:
PASS = normal
FAIL = consult Milltronics.
P911 mA Output Test
Tests the mA output value for the output selected, outputs 1 to 3 (P911 01 to -03)
Displays the value from the previous measurement. A test value can be entered and
the displayed value is transmitted to the output. Upon returning to the run mode, the
parameter assumes the actual mA output level. (f=0)
Use the up and down to scroll the value
P914 mA Input Value
Displays the current mA input value for the input selected, inputs 1 to 2 (P914 01 to
-02).
Note:
Not applicable if mA I/O board is not connected.
P918 Speed Input Frequency
Displays the frequency of the speed input signal in Hertz.
P940 Load Cell mV Signal Test
Displays the raw (unbalanced) mV signal input for the selected load cell, load cells A
to D* (P940 01 to 04)
Range 0.00 - 60.00 mV.
*depending on the number of load cells selected by P003, not all load cells may be in
use.