Maintaining Your Analyzer
10-8
At the “Maintenance” screen, SELECT “Lyse Prime” to prime the tubing with lyse and the
priming progress will be displayed at the bottom of the screen, as Figure10-5 shows. When
the priming is done, the screen will display “Lyse Prime End”.
Figure10-5 Lyse priming
10.3.4 Zap Apertures
You can perform the “Zap Aperture” procedure to unclog the apertures or prevent clogging.
At the “Maintenance” screen, SELECT “Zap Aperture” to zap the apertures and the zapping
progress will be displayed at the bottom of the screen, as Figure10-6 shows. When the
zapping is done, the screen will display “Zap Apertures End”.