BeneLink Module Operator’s Manual 4-79
Monitor output: alarms from Newport E360 Ventilator
Priority Label
High High
Technical
alarms
Fan Failure
Air Flow Sensor EEPROM Failure
O2 Flow Sensor EEPROM Failure
Medium
O2 and air supply
O2 Sensor Unconnected
Medium
Technical
alarms
If the monitor shows the alarm [Medium Technical
alarms], the external device has one or more of the
following alarms:
Flow Sensor Cal Failed
FiO2 Sensor Bad
O2 Sensor Cal Failed
External Battery
Check Flow Sensor Board
NO TEST
Low
Lo
w
Technical
alarms
If the monitor shows the alarm [Low Technical
alarms], the external device has one or more of the
following alarms:
I:E Ratio Inverse violation
Plimit<Pbase
Psupport+Pbase>60cmH
2
O
Pbase>Low Paw
Tidal Volume Out of Range
Flow Out of Range
Ti Out of Range
Rate Out of Range
Psupport Out of Range
Plimit Out of Range
PEEP/CPAP Out of Range