EasyManuals Logo

Mindray M8 User Manual

Mindray M8
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
1-4 Safety Precautions
1.3.1 Electric Safety
WARNING:
1.
Connect the power plug of this system and power plugs of the
peripherals to wall receptacles that meet the ratings indicated
on the rating nameplate. Using a multifunctional receptacle may
affect the system grounding performance, and cause the
leakage current to exceed safety requirements. Use the power
cord accompanied with the system provided by Mindray.
2.
Disconnect the AC power before you clean or uninstall the
ultrasound machine, otherwise, electric shock may result.
3.
When using peripherals not powered by the auxiliary output of the
ultrasound system, or using peripherals other than permitted by
Mindray, make sure the overall leakage current of peripherals and
the ultrasound system meets the requirement of the local medical
device electrical regulation (like enclosure leakage current should
be no more than 500uA of IEC 60601-1-1), and the responsibility is
held by the user.
4.
In maintenance or assembly/disassembly, make sure other
cables are connected well before the battery connecting cable is
connected, otherwise the system may be damaged due to
hot-plug.
5.
Do not use this system simultaneously with equipment such as
an electrosurgical unit, high-frequency therapy equipment, or a
defibrillator, etc.; otherwise electric shock may result.
6.
This system is not water-proof. If any water is sprayed on or into
the system, electric shock may result.
CAUTION:
1.
DO NOT connect or disconnect the systems power cord or its
accessories (e.g., a printer or a recorder) without turning OFF
the power first. This may damage the system and its
accessories or cause electric shock.
2.
Avoid electromagnetic radiation when perform performance test
on the ultrasound system.
3.
In an electrostatic sensitive environment, don’t touch the device
directly. Please wear electrostatic protecting gloves if
necessary.
4.
You should use the ECG leads provided with the ECG module.
Otherwise it may result in electric shock.
5. Maximum output power
of the trolley is 240 VA.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M8 and is the answer not in the manual?

Mindray M8 Specifications

General IconGeneral
BrandMindray
ModelM8
CategoryMedical Equipment
LanguageEnglish

Related product manuals