11-6 DICOM/HL7 
DICOM storage preset items are described as follows: 
After setting the servers in the DICOM local screen, the 
names will appear in the drop-down list. Select the name of 
the storage server. 
The default is xxx-Storage, user-changeable. 
Application Entity title. It should be consistent with that of 
the storage server. 
DICOM communication port, 104 is the default. The port 
should be consistent with that of the storage server port. 
Select whether to encrypt the data during network 
transportation.   
Refers to the amount of time after which the system will 
stop trying to establish a connection to the service. 
Select the Compression mode: uncompressed, RLE, JPEG 
and JPEG2000. 
Select the JPEG Compression ratio: lossless, low, medium 
and high. 
If SCP supports this function, select it. 
Set the frame range for transferring cine files to DCM multi-
frame files. 
Select structured report sending options. 
Select whether to encapsulate PDF format reports in 
DICOM standard. 
Set the storage mode for image and cine file: 
Parallel file: save the current file, and is ready for the 
storage of the next file. 
Parallel frame: send the current frame, and is ready for 
sending the next frame. 
Files of images that are saved in DCM format through 
DICOM contain transducer serial number information.   
Add the DICOM service to the service list. 
Select to cancel parameter setting. 
Select an item in the service list, change the parameters in 
the above area, and tap [Update] to update the item in the 
service list. 
Select to delete the selected service from the service list. 
Select an item in the service list. Tap [Default] and you will 
see “Y” in the Default column.