1-8 Safety Precautions 
Please read the following precautions carefully to ensure the safety of the patient and the operator 
when using the probes. 
The ultrasonic probe is only for use with the specified ultrasonic 
diagnostic system. Please refer to the 2.3.2 Probes and Needle-
guided Brackets Available to select the proper probe. 
Confirm  that  the  probe  and  cable  are  normal  before  and  after 
each examination. A defective probe may cause electric shock 
to the patient. 
Do not subject the probe to shock. A defective probe may cause 
electric shock to the patient. 
Do not disassemble the probe as this may cause electric shock 
or malfunction. 
Never immerse the probe connector into liquids such as water 
or  disinfectant  because  the  connector  is  not  waterproof. 
Immersion may cause electric shock or malfunction. 
A  probe  sheath  must  be  installed  over  the  probe  before 
performing intra-cavity or intra-operative examination. 
 
When using the probe, wear sterile gloves to prevent infection. 
Be sure to use sterile ultrasound gel. Please use the ultrasound 
gel compliant with the relevant local regulations. And manage 
the ultrasound gel properly to ensure that it does not become a 
source of infection. 
In normal diagnostic ultrasound mode, there is no danger of a 
normal-temperature  burn;  however,  keeping  the  probe  on  the 
same  region  of  the  patient  for  a  long  time  may  cause  such  a 
burn. 
Do not use the carrying case for storing the probe. If the carrying 
case is used for storage, it may become a source of infection. 
It  is  required  to  practice  ALARA  when  operating  ultrasound 
system. Minimize the acoustic power without compromising the 
quality of images. 
The probe and the accessories supplied with it are not delivered 
disinfected or sterilized. Sterilization (or high-level disinfection) 
is required before use. 
Disposable  components  (for  example  the  probe  sheath,  the 
sterile gloves) are packaged sterile and are single-use only. Do 
not use if the integrity of the packaging has been violated or if 
the  expiration  date  has  passed.  Use  disposable  components 
which comply with the relevant local regulations. 
Please  use  the  disinfection  or  sterilization  solution  that 
recommended in this operator’s manual; otherwise Mindray will 
not be liable for damage caused by other solutions. If you have 
any  questions,  please  contact  Mindray  Customer  Service 
Department or sales representative.