Probes and Biopsy 13-35
4. Hold the transducer. Press the biopsy needle pressure position to separate it from needle
guided V-shaped block.
5. Put the needle into the needle guided-bracket, and the needle leans to V-shaped block.
6. Hold the transducer, and release the pressure position of the needle. Adjust the needle-type
adjusting nut manually (following the direction of the arrow). The needle moves smoothly at
the vertical direction due to its gravity.
CAUTION:
Ensure that all guide parts are properly fixed prior to performing a
biopsy.