Alarm Watch Alarm Watch
23 - 4 V Series Operating Instructions
3. Select the Add Bed button, then select a bed from the list box.
The bed displays in the Alarm Watch Setup grid.
NOTE: Only locations where room/bed information was entered are available
in the list.
NOTE: The beds are listed alphabetically in the list box.
NOTE: One Host monitor can watch a maximum of 16 beds.
4. Select the Cancel or Accept button.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.
23.1.4 Removing a Bed
1. Select the A-W tile.
The Alarm Watch Setup dialog displays.
2. Select an unlocked bed in the Alarm Watch Setup grid.
NOTE: Locked beds (indicated by the symbol) cannot be removed from the
Alarm Watch Setup grid. Refer to "Removing Alarms from the Watched
Alarms List" on page 23-5 for additional information.
3. Select the Remove Bed button.
4. Select the Cancel or Accept button.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.
23.1.5 Adding Alarms to Watched Alarm List
NOTE: Alarms added in this dialog are not saved to the default System Alarm
Watch locations.
1. Select the A-W tile.
The Alarm Watch Setup dialog displays.
2. Select a bed in the Alarm Watch Setup grid.
3. Select the Alarms button.
4. Select an alarm from the Choices list (left side of dialog), then select the Add button.
The alarm moves to the Watched Alarms list (right side of dialog).
5. Continue to select the Cancel or Accept button to return to the Main display.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.