System Dialog System Setup
26 - 6 V Series Operating Instructions
FIGURE 26-3 Example System Dialog (Alarms Tab)
4. Select the buttons described in TABLE 26-3 to configure the tab.
TABLE 26-3
BUTTON DESCRIPTION / COMMENTS CHOICES*
Audio Pause Timeout Allows an authorized user to choose which
settings are available in the Alarm Setup
dialog Audio Pause functions. This setting
controls for how long an alarm is muted.
1, 2, 5,10 min, Permanent
1, 2, 5, 10 min
1, 2, 5 min
1,2 min
NOTE: The Audio Pause configuration applies to all
patient sizes.
Lethal Alarms Allows an authorized user to determine
whether or not a user can modify the lethal
arrhythmia settings and decide whether
Asystole, VTach and VFib alarms can be
turned off by the user.
Locked
Unlocked
Apnea Alarms Allows an authorized user to enable Apnea
alarms at the monitor.
Locked
Unlocked
Latch Alarms Configures whether or not to latch the
physiological alarms.
All On
All Off
Lethal Only
Alarm/Response
Order
Configures the order of the Alarm Limit
dialog and Alarm Responses dialog tabs.
Refer to "Configuring the Alarm/
Response Tab Order" on page 26-
7.
Mixed Agent Audio Allows an authorized user to enable or
disable the Mixed Agent Audio.
On
Off
* The factory default configuration settings are in bold.

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals