V Series Operating Instructions 26 - 7
System Setup System Dialog
5. Select the Cancel or Accept button.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.
26.2.5 Configuring the Alarm/Response Tab Order
Define the order in which the tabs display in the Alarm Limits and Responses tab in the Alarm Limit/
Response Tab Order dialog. The factory default alarm tab order is ECG/Resp, Pressure, Temp, Gases,
CCO/Hemo, and Oximetry.
To configure the tab order:
1. Select the Setup tab from the Navigation Area.
2. Select the System button, enter the password, then select the Accept button.
3. Select the Alarms tab.
4. Select the Alarm/Response Order button.
The Alarm Limit/Response Tab Order dialog displays (as shown in FIGURE 26-4).
NOTE: The Mixed Agent Audio configuration
applies to all patient sizes.
ECG Lead Fault Audio Configures the audio alert when there is a
lead fault which impacts HR or Arrhythmia.
Informational - Single Tone
Informational - Five Tones
Informational - Continuous
Low Priority
Medium Priority
High Priority
SpO2 Sensor Audio Configures the audio alert when there is no
SpO2 sensor, or the SpO2 Sensor off
condition occurs.
Off
Informational - Single Tone
Informational - Five Tones
Informational - Continuous
Low Priority
Medium Priority
High Priority
Minimum Alarm
Volume
Allows an authorized user to configure the
minimum alarm volume.
1, 2, 3, and 4
Alarm Setup Password Allows an authorized user to change the
Alarm Setup password
N/A
TABLE 26-3
BUTTON DESCRIPTION / COMMENTS CHOICES*
* The factory default configuration settings are in bold.