System Dialog System Setup
26 - 20 V Series Operating Instructions
26.2.12 Removing Alarms from the Watch Alarm List
1. Select the Setup tab from the Navigation Area.
2. Select the System button, enter the password, then select the Accept button.
3. Select the Alarm Watch tab.
4. Select the Alarms button.
5. Select an unlocked alarm from the Watched Alarms list (right side of dialog), then select the
Remove button.
The alarm moves to the Choice list (left side of dialog).
NOTE: Asystole, V-Fibrillation, and V-Tachycardia are automatically included
in the watched alarms list. They are indicated by the locked icon
(meaning they cannot be removed from the list).
6. Continue to select the Cancel or Accept button to return to the Main display.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.
26.2.13 Alarm Watch Response
The Alarm Watch Response is Visual/Audio. This setting indicates that when a configured alarm
occurs, a visual indication displays and an audio alarm enunciates.
26.2.14 Configuring the System Tab
Select the System tab to configure the demo mode and service settings. These functions apply to all
patient sizes.
To configure the System tab:
1. Select the Setup tab from the Navigation Area.
2. Select the System button, enter the password, then select the Accept button.
3. Select the System tab.
The System tab displays (as shown in FIGURE 26-12).