V Series Operating Instructions 26 - 21
System Setup System Dialog
FIGURE 26-12 Example System Dialog (System Tab)
4. Select the buttons described in TABLE 26-11 to configure the tab.
TABLE 26-11
BUTTON/
CHECK BOX DESCRIPTION / COMMENTS CHOICES*
Demo Mode Allows an authorized user to choose a
demonstration mode for in-servicing staff or
testing product features.
Enable, Disable
Demo Scenario Allows an authorized user various
demonstration scenarios that are intended to
non-clinical conditions.
List all scenarios contained in the
Scenarios folder.
Alarm Acknowledge
Source
This checkbox is related to the Remote View
feature. When enabled, allows the alarms at
the source location to be silenced by
selecting a button from the remote location.
On, Off
Restore Factory
Defaults
Allows an authorized user to reestablish the
original database power up settings to
factory default values.
N/A
NOTE: Refer to "Restore Factory Defaults" on page 26-22 for a
list of exceptions.
Change Password Allows an authorized user to change the
System password.
N/A
Service Allows an authorized user access to the
password protected Service Diagnostic
mode.
N/A
Export Config Copies the current settings to the external
drive.
N/A
* The factory default configuration settings are in bold.