8.1.9 Upgrading Nurse Call Function and DIAP Protocol Port
Remove the power management board, and the stopple of multifunctional connector on
the rear panel of the equipment (if there is one) as described in 6.3 Disassembling the Main
Unit. Then assemble the power management board with multifunctional connector. Then
reassemble the main unit.
Once the above functions are successfully upgraded, perform related tests on the
equipment as described in 4.4 Module Performance Tests, 4.6 Touchscreen Calibration,
and 4.7 Recorder check.
8.2 Software Upgrade
You can upgrade system software and module software by installing and running the
Mindray Patient Monitor Software Upgrade Tool on a PC with Windows operating system.
Connect the monitor to be upgraded and a PC running the upgrade tool to the same
network, or directly connect the monitor and the PC via a crossover network cable. Then
configure the IP address of the PC. You can upgrade the following software:
No. Type PN Description
1
System software
package
110-002646-00 System software package
2
Module software
110-000889-00 Mindray NIBP function software
3 110-001842-00 Mindray SpO
2
function software
4 110-002027-00 Temp function software
5
Power Management
Software
110-002525-00 Power Management Software
6 MEWS
115-030317-00
MEWS
Note: No specific sequence is required for the upgrading of above software. For detailed
information, please refer to 8.2.2Software Upgrade Procedure.
NOTE
The software upgrading could result in clearing the historic patient data. It is
recommended to export patient data before software upgrading.