6-2
6.3 Disassembling the Main Unit
NOTE
The recorder can be disassembled separately.
To disassemble the equipment, place the equipment on a work surface free from
foreign material, avoiding damaging the antiglare screen, touchscreen and the
knob. Be careful not to break the two cotters on the front ends of rear housing.
All the operations should be performed by qualified service personnel only. Make
sure to put on the insulating gloves during service operations.
Operations relating to optional parts may not apply to your equipment.
6.3.1 Disassembling the SmarTemp
TM
Module (Optional)
1. Lay the monitor on a table as shown below. Unscrew the three M3×6 screws, pull the
SmarTemp
TM
module up, and disconnect the Temp cable.
6.3.2 Disassembling the Infrared Ear Thermometer Module (Optional)
Lay the monitor on a table as shown below. Unscrew the three M3×6 screws, pull the
Infrared Ear Thermometer module up, and disconnect the infrared ear temperature cable.
T
emp
cable
Infrared ear
temperature
cable