5-6 
 
5.6.9 Data Storage Failure 
Symptoms  Possible Cause  Troubleshooting 
Fails to review 
archived patient 
data 
Abnormal patient admitting Admit the patient properly. 
SD card full; unavailable for 
more patient data 
Delete garbage patient data, remove the 
related alarm, and readmit the patient. 
The main board failed.  Replace the main board. 
SD card failure 
SD card not formatted  Format the SD card. 
SD card failure  Replace the SD card. 
SD card is locked  Unlock the SD card. 
Main board defective  Replace the main board. 
 
 
5.6.10 Wired Network Related Problems 
Symptoms  Possible Cause  Troubleshooting 
Unable to connect 
the wired network 
Incorrect LAN cable 
connection 
Check LAN cable connection. LAN cable shall not 
be longer than 50 m. 
Incorrect IP address 
configuration 
Check for IP address conflict. If yes, reconfigure 
the IP address. 
Cable defective 
1. Check that the cable between the power 
management board and main board is correctly 
connected. 
2. Check that the cable between power 
management board and interface board is 
correctly connected. 
The connector board 
failed. 
Replace the connector board. 
Power management 
board defective 
Replace the power management board. 
The main board 
failed. 
Replace the main board. 
The monitor is 
frequently off line or 
disconnects from 
the network. 
Incorrect LAN cable 
connection 
Check LAN cable connection. LAN cable shall not 
be longer than 50 m. 
The monitor is 
connected to a LAN 
but cannot view 
other patients 
under the View 
Others mode 
Incorrect LAN cable 
connection 
Check LAN cable connection. LAN cable shall not 
be longer than 50 m. 
Excessive requests 
for viewing the 
monitor at the same 
time 
A monitor can only be viewed by 4 other 
monitors at the same time under the View Others 
mode. The excessive view requests system will be 
ignored.