6.4.9 Removing the Alarm Lamp
Unscrew the four PT2×6 screws indicated below and take out the alarm lamp board and
alarm lamp.
6.5 Disassembling the SmarTemp
TM
Module (Optional)
6.5.1 Removing the SmarTemp
TM
Module PCBA and SmarTemp
TM
Module
Power Board PCBA
Unscrew the two M3×6 screws and the two M3 screws. Remove the cover board and metal
sheet. Disconnect the Temp board cable and the cable between Temp isolation power board
and Temp board. Unscrew the four M3×6 screws, you can take out the SmarTemp
TM
module
PCBA. Then unscrew the three M3×6 screws, you can take out the SmarTemp
TM
module
power board PCBA.
Cover board
Metal sheet
Screw for cover board
Temp board cable
Cable between the
isolation power
board and Temp
board
4 screws
Sma
emp
TM
module PCBA
Screw for metal sheet
Sma
emp
TM
module power
board PCBA
3 screws