1-2
1.2.1 Dangers
There are no dangers that refer to the product in general. Specific “Danger” statements may
be given in the respective sections of this manual.
1.2.2 Warnings
WARNING
All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel.
There is high voltage inside the equipment. Never disassemble the equipment
before it is disconnected from the AC power source or the battery.
When you disassemble/reassemble a parameter module, a patient leakage current
test must be performed before it is used again for monitoring.
The equipment must be connected to a properly installed power outlet with
protective earth contacts only. If the installation does not provide for a protective
earth conductor, disconnect it from the power line and operate it on battery power,
if possible.
Disposal of the packaging material should observe the applicable waste control
regulations and keeping it out of children’s reach.
1.2.3 Cautions
CAUTION
Make sure that no electromagnetic radiation interferes with the performance of the
equipment when preparing to carry out performance tests. Mobile phone, X-ray
equipment or MRI devices are a possible source of interference as they may emit
higher levels of electromagnetic radiation.
Before connecting the receiver to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the unit’s
label or in this manual.
Protect the equipment from damage caused by drop, impact, strong vibration or
other mechanical force during servicing.
1.2.4 Notes
NOTE
Refer to Operator's Manual for detailed operation and other information.